PulseLoop support: ๐Ÿงช Limited โ€” the protocol is confirmed on hardware, this ring isn't

The TK5 pairs with the SmartHealth app (com.zhuoting.healthyucheng) and speaks the Yucheng YCBT protocol on a be940 service โ€” nothing in common at the wire level with the 56ff / Jring or Colmi / Yawell QRing families.

Limited support โ€” the protocol is proven, the TK5 hasn't been re-tested

The YCBT stack is confirmed working on real hardware โ€” but on a sibling ring, the Colmi R09, not on a TK5. Pairing, the handshake, history sync and live measurements all check out there, and both rings run the identical driver.

What's still open here is TK5-specific: a few scales and payloads (blood sugar, temperature, find-device) need one confirmed reading โ€” see Needs on-device confirmation. Every decoded metric is range-gated before storage, so a misdecode is dropped rather than saved as garbage.

Not the only ring that speaks it

The TK5 is one of three ring families PulseLoop drives over YCBT. The others are the R10M / LittleMeatball and Colmi rings that ship with SmartHealth instead of QRing. The protocol is byte-identical, so they share the whole driver (the device-neutral YCBT* core); each family adds only a coordinator with its advertisement matcher and capability set, plus a small profile for firmware quirks. They differ in three ways:

TK5 R10M SmartHealth-Colmi
Advertisement TK5 <4 hex> โ€” unambiguous, so it auto-detects R10M <4 hex> and the be940000 service a Colmi-line name, which a QRing Colmi can also carry, so PulseLoop asks which app the ring came with
Hardware validation none โ€” the protocol is proven on a sibling, this ring isn't โœ… full session on FW 2.32 an R99 runs it daily
SupportFunction bitmap (02 01) gates temperature, BP, stress, fatigue, blood sugar. HRV is not gated โ€” it was observed working on this ring gates all of those and HRV gates those and HRV, which the tested R09 denies

A fix to any YCBT* file fixes all three rings; a regression in one breaks all three. The R10M also suppresses two commands the other two send โ€” see its firmware quirks.

At a glance

Detail
SoC JieLi (chip scheme 3/4/5 โ€” inferred from the AE00 RCSP service; exact part โ“)
Bluetooth BLE (version โ“)
PPG sensor โ“ Unknown โ€” green LED for HR, red/IR for SpOโ‚‚
Accelerometer Yes (steps decoded; part โ“)
Battery / life โ“ Unknown
Waterproof โ“ Unknown
Price โ“ Unknown
Protocol Yucheng YCBT โ€” variable-length frames, CRC16/CCITT-FALSE, cleartext
App SmartHealth (com.zhuoting.healthyucheng)
Advertised name TK5 <4 hex> โ€” e.g. TK5 24AA
Custom firmware โ“ Unknown

Protocol

Property Value
Service be940000-7333-be46-b7ae-689e71722bd5
Command char be940001 โ€” write and indicate (the app writes here and gets replies here)
Stream char be940003 โ€” indicate: live vitals and all history data frames
Frame [type:1][cmd:1][len:2 LE][payload:N][crc16:2 LE], len = total frame length
CRC CRC16/CCITT-FALSE (poly 0x1021, init 0xFFFF, no reflection), little-endian
Epoch Seconds since 2000-01-01, in the ring's local wall-clock (no timezone concept)
Fragmentation Frames split across notifications at MTUโˆ’3; reassembled by the declared length
History 05 <key> query โ†’ header โ†’ concatenated data frames โ†’ 05 80 terminal โ†’ mandatory 05 80 {00} ACK
Encryption None on the health protocol. The AE00 service is JieLi RCSP โ€” see below

Capabilities

Everything here is decoded from the vendor SDK and covered by unit tests against fixture bytes. The right-hand column is the honest one: what a physical ring still has to confirm.

Ring-declared vs. baseline. The rows marked ๐Ÿ”“ ring-declared are no longer promised by the app at all: they are offered only if this unit's 02 01 capability bitmap sets their bit (YCBTSupportFunction โ†’ TK5Coordinator.bitmapGatedCapabilities). Everything else is a baseline promise โ€” the app claims it unconditionally, and the bitmap can only ever add, never remove.

Capability Status Needs on-device confirmation Notes
Heart rate โ€” spot ๐Ÿงช โ€” 03 2f mode 00 โ†’ 06 01 stream; the standard 180D char is deliberately ignored (see below)
Heart rate โ€” live ๐Ÿงช โ€” verified against the capture (climbed 82โ†’86)
Heart rate โ€” history ๐Ÿงช โ€” 05 06 query, 6-byte records
SpOโ‚‚ โ€” spot ๐Ÿงช โ€” red/IR LED, 03 2f mode 02
SpOโ‚‚ โ€” history ๐Ÿงช โ€” dedicated 05 1A log and the All record. Not separately gated: no bit names it โ€” it is one of the SpOโ‚‚ sources ISHASBLOODOXYGEN already grants
Steps / distance / calories ๐Ÿงช distance & calories byte offsets in the 06 00 live push live push + 05 02 history buckets + the All record's cumulative counter
Sleep (light / deep / awake) ๐Ÿงช โ€” 05 04 timeline; stage = tag & 0x0F
REM sleep ๐Ÿงช โ€” stage tag 3 โ€” a stage inside the timeline ISHASSLEEP grants, so no bit names it and it is not gated
HRV ๐Ÿงช the three tail bytes of the 01 45 monitor enable spot (03 2f mode 0a) + the All and body-data records. Baseline โ€” observed on the ring (see above)
Blood pressure โ€” spot ๐Ÿ”“ per-mode stop (03 2f {00, 01}) ring-declared (ISHASTESTBLOOD, byte 15 bit 2). 03 2f mode 01; no cuff calibration, so treat as a trend, not a number
Blood pressure โ€” history ๐Ÿ”“ โ€” ring-declared (ISHASBLOOD, byte 0 bit 0). Dedicated 05 08 log + the All record
Skin temperature ๐Ÿ”“ one real reading (the string-concat scale) ring-declared (ISHASTEMP, byte 8 bit 0). Dedicated 05 1E log + the All record; monitor enabled by 01 20
Stress ๐Ÿ”“ โ€” ring-declared (IS_HAS_PRESSURE, byte 22 bit 6). Ring-stored, from the body-data record (05 33) โ€” the SDK's pressure field
Fatigue ๐Ÿ”“ โ€” ring-declared on the same bit as stress: no ISHASFATIGUE exists, but the vendor app gates the whole 05 33 query on IS_HAS_PRESSURE, and fatigue (the SDK's body field) is a field of that one record. No body data โ‡’ neither score
Blood sugar ๐Ÿ”“ one real reading โ€” the scale is inferred, not observed ring-declared (ISHASBLOODSUGAR, byte 17 bit 3). All record @17 and the comprehensive record (05 2F); tenths of mmol/L โ†’ mg/dL
Respiratory rate ๐Ÿงช โ€” All record @10
VOโ‚‚max ๐Ÿงช raw byte taken unscaled body-data record @16
Battery level ๐Ÿงช โ€” in-band: 02 00 reply payload[5], plus the unprompted 06 15 push
Find device ๐Ÿงช the command's three payload bytes 03 00 {01, 05, 02} โ€” the app's literal arguments; the SDK never names them
Measurement intervals ๐Ÿงช โ€” five 01 xx {enable, interval} monitors; floored at the firmware's 30-min minimum
Periodic re-sync while connected โœ… โ€” every 30 min (SmartHealth's own cadence), plus a post-workout vitals pass
Continuous background sync โŒ โ€” the ring is only read while connected
FW update via app โŒ โ€” out of scope โ€” would require the JieLi RCSP auth (see below)

Needs on-device confirmation

The open items, in priority order. Each is range-gated in code, so a wrong guess degrades to "no reading", never to a wrong reading:

  1. Blood-sugar scale โ€” the tenths-of-mmol/L reading is inferred from SmartHealth's database column and chart filter, not from an observed non-zero record. Cross-check one reading.
  2. Temperature scale โ€” the int/fraction pair is string-concatenated (5 โ†’ .5, 25 โ†’ .25), not divided. Confirm a real reading lands at ~36.x ยฐC.
  3. Find-device payload โ€” {1, 5, 2} replays the app's own button; the SDK never names the arguments. First suspect if the ring doesn't buzz.
  4. Per-mode stop โ€” an SpOโ‚‚/HRV sweep is stopped with its own mode byte. If the LED stays on afterwards, this is the line to look at.
  5. HRV monitor tail bytes โ€” 01 45 {enable, interval, 0, 0, 0}; only the first two arguments are named in the SDK, so the rest are zero-filled rather than guessed.
  6. Sleep โ€” the multi-session timeline is decoded from the SDK and unit-tested, but no YCBT ring has been worn overnight yet.

The AE00 service

The ring exposes a second service, AE00, carrying an encrypted FE DC BA โ€ฆ / 02 "pass" handshake. It looks like a login gating the health data. It is not โ€” and that is now settled on hardware, not just from the SDK.

It is JieLi RCSP โ€” the chipset vendor's challenge-response auth, an entirely separate subsystem from the Yucheng health protocol. It authorizes the JieLi feature set only: OTA / firmware update, watch-face upload, and log extraction. The health commands on be940001 are plaintext, CRC-framed and carry no auth of any kind. The AES key lives in a native library (libjl_rcsp.so) and is not recoverable from the decompile.

This was the project's one potential hard stop: the SDK proves the two code paths are independent, but it cannot prove a given firmware doesn't refuse health commands until RCSP auth completes. The Colmi R09 answered that โ€” it reports chip scheme 4 (JieLi) and every health command, history query and measurement still answered in plaintext, with no AE00 exchange at all. PulseLoop implements none of RCSP, and doesn't need to.

How PulseLoop diverges from SmartHealth

We never delete records off the ring. SmartHealth issues the Health-Delete opcodes (05 40โ€ฆ4E) after a sync, so its next sync is small. PulseLoop doesn't: a delete that races a failed decode destroys data that exists nowhere else, and it makes the vendor app and PulseLoop mutually destructive on the same ring.

The consequence is that the ring replays its entire log on every sync, so deduplication is app-side: every history sample upserts on (kind, timestamp), activity buckets upsert by start epoch, and the cumulative step counter is a per-day max ratchet. All three are idempotent under replay, so a double-sync produces no duplicates. The cost is a longer sync as the ring's log fills.

Known limitations

  • ~8-day history horizon. RingEventBridge drops any history sample, sleep session or activity timestamp outside now โˆ’ 8 days โ€ฆ now + 1 hour. A ring's log can hold records stamped under a previous clock, which decode hours or days out of place โ€” and because history rows upsert, one misdecoded record wouldn't merely flicker, it would re-persist on every future sync. Records older than the window are therefore not imported.
  • No background sync while disconnected. The ring keeps logging on its own schedule (that's what the 01 xx monitors are for), but PulseLoop only reads it while connected: on connect, every 30 minutes thereafter, and after a workout.
  • No firmware updates or watch faces. Both live behind the JieLi RCSP auth (see above) and are out of scope.
  • Standard heart-rate service is deliberately ignored. The TK5 exposes 180D/2A37, but it emits a cached resting HR (~87 bpm) even off-finger, which would mask real readings. Live HR comes solely from the proprietary 06 01 stream, as in the official app.
  • Timestamps are timezone-naive. The ring stores local wall-clock seconds with no timezone byte, so the decoder un-applies the device's UTC offset to recover the true instant. The offset is resolved at each record's own date, so history read across a DST change decodes correctly; the only residue is the ambiguous hour a fall-back repeats and the hour a spring-forward skips, which the wire format simply cannot disambiguate.

See the SmartHealth-app Colmi rings that share this driver, or the hardware overview for the cross-manufacturer comparison.